Medonica Healthcare Pvt. Ltd.
Medonica Healthcare Pvt. Ltd.
Request a quote

Medonica Count 60 Touch Screen 3-Part Differential Haematology Analyzer

Three-part differential haematology analyzer with 21 parameters, 10.4-inch touch screen and throughput of 60 samples per hour.

  • 21 parameters and 3 histograms
  • 10.4-inch touch screen
  • 60 samples per hour
  • Stores 100,000 results with histograms
  • LIS/HIS and HL7 support; closed system with RFID
Add to enquiry

For in vitro diagnostic use by laboratory and healthcare professionals only. Please refer to the instructions for use supplied with each product for complete protocol and performance details.

Overview

The Medonica Count 60 performs a complete blood count with three-part differentiation of white blood cells, reporting 21 parameters and 3 histograms on a 10.4-inch touch screen. It stores 100,000 patient results with histograms, supports LIS/HIS and HL7, and uses a closed reagent system with RFID detection.

Intended use: Complete blood count with three-part differential of white blood cells on whole blood or pre-diluted samples.

Standards: EN 61010-1:2010, EN 61010-2-101 (electrical safety, as stated by the manufacturer)

Specifications

Specifications for Medonica Count 60 Touch Screen 3-Part Differential Haematology Analyzer
Power / input ratingAC 100–240 V, 50/60 Hz, 180 W
Dimensions (L × W × H)330 × 440 × 500 mm
LinearityWBC 0–10.0 × 10⁹/L (≤ ±0.3), 10.1–99.9 × 10⁹/L (≤ 5 %); RBC 0–1.00 × 10¹²/L (≤ ±0.05), 1.01–9.99 × 10¹²/L (≤ 5 %); HGB 0–70 g/L (≤ ±2 g/L), 71–300 g/L (≤ 2 %); PLT 0–100 × 10⁹/L (≤ ±10), 101–999 × 10⁹/L (≤ 10 %)
PrincipleElectrical impedance (WBC, RBC, PLT); photoelectric colorimetry, cyanide-free (HGB)
Parameters21 including WBC, LY#, MID#, GRAN#, LY%, MID%, GRAN%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW, PCT, P-LCR, P-LCC
Display10.4-inch colour LCD touch screen
Interfaces4 USB ports; internal 57 mm thermal printer; parallel port
Data storage100,000 results with histograms
Aspiration volumeWhole blood 10 µL; pre-diluted 20 µL
Analysis modeWhole blood; pre-diluted
Throughput60 samples per hour
Aperture diameterWBC 100 µm; RBC/PLT 68 µm
Quality control and calibrationL-J, X-R, X-B and X QC; manual and automatic calibration
Dilution ratioWBC/HGB: whole blood 1:232, capillary 1:400; RBC/PLT: whole blood 1:40,000, capillary 1:45,000