HbQue Haemoglobin and HCT Meter
Handheld reflectance photometer for haemoglobin and haematocrit from 10 µL of capillary or venous whole blood.
Three-part differential haematology analyzer with 21 parameters, 10.4-inch touch screen and throughput of 60 samples per hour.
For in vitro diagnostic use by laboratory and healthcare professionals only. Please refer to the instructions for use supplied with each product for complete protocol and performance details.
The Medonica Count 60 performs a complete blood count with three-part differentiation of white blood cells, reporting 21 parameters and 3 histograms on a 10.4-inch touch screen. It stores 100,000 patient results with histograms, supports LIS/HIS and HL7, and uses a closed reagent system with RFID detection.
Intended use: Complete blood count with three-part differential of white blood cells on whole blood or pre-diluted samples.
Standards: EN 61010-1:2010, EN 61010-2-101 (electrical safety, as stated by the manufacturer)
| Power / input rating | AC 100–240 V, 50/60 Hz, 180 W |
|---|---|
| Dimensions (L × W × H) | 330 × 440 × 500 mm |
| Linearity | WBC 0–10.0 × 10⁹/L (≤ ±0.3), 10.1–99.9 × 10⁹/L (≤ 5 %); RBC 0–1.00 × 10¹²/L (≤ ±0.05), 1.01–9.99 × 10¹²/L (≤ 5 %); HGB 0–70 g/L (≤ ±2 g/L), 71–300 g/L (≤ 2 %); PLT 0–100 × 10⁹/L (≤ ±10), 101–999 × 10⁹/L (≤ 10 %) |
| Principle | Electrical impedance (WBC, RBC, PLT); photoelectric colorimetry, cyanide-free (HGB) |
| Parameters | 21 including WBC, LY#, MID#, GRAN#, LY%, MID%, GRAN%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW, PCT, P-LCR, P-LCC |
| Display | 10.4-inch colour LCD touch screen |
| Interfaces | 4 USB ports; internal 57 mm thermal printer; parallel port |
| Data storage | 100,000 results with histograms |
| Aspiration volume | Whole blood 10 µL; pre-diluted 20 µL |
| Analysis mode | Whole blood; pre-diluted |
| Throughput | 60 samples per hour |
| Aperture diameter | WBC 100 µm; RBC/PLT 68 µm |
| Quality control and calibration | L-J, X-R, X-B and X QC; manual and automatic calibration |
| Dilution ratio | WBC/HGB: whole blood 1:232, capillary 1:400; RBC/PLT: whole blood 1:40,000, capillary 1:45,000 |
Handheld reflectance photometer for haemoglobin and haematocrit from 10 µL of capillary or venous whole blood.
Compact three-part haematology analyzer with a 10.4-inch touch screen, 9 µL sample and 60 samples per hour.
Three-part differential haematology analyzer with 22 parameters, a double chamber, 8-inch touch screen and cyanide-free haemoglobin.